Description
- The wrenches are not sterilized prior to shipping.
- They must be cleaned, disinfected, and sterilized before use.
- When cleaning, do not use abrasive detergents and do not apply excessive pressure.
- To avoid damage do not clean with a wire brush or metal pads.
- Use only cleaning and disinfectant solutions with a ph between 5 and 10.
- Please follow the cleaning/disinfectant manufacturers’ instructions (e.g. intended use, dosage, time, renewal period).
- Frequent but careful cleaning has little effect on the life of the ratchet.
- Product life is determined by wear.
- Recalibrate the wrench once a year. The wrench must be cleaned and sterilized after service.
Recommended Sterilization Method
- Loosen the ratchet completely prior to sterilization
- Method Fractionated pre-vacuum process (according to ISO 17665)
- Temperature Heat up to 134⁰C (137⁰C Maximum)
- Pressure 3 pre-vacuum phases with at least 60 millibar pressure
- Holding Time at least 5 minutes
- Drying Time at least 10 minutes
- After sterilization checks the packaging of the sterilized instruments for damage. Check the sterilization indicators.
Storage
- Store the sterilized wrenches in a dry, dust-free, corrosive vapor-free area at a moderate temperature.